常见问题 2026-03-05 关键词:医疗设备,医疗影像设备,设备维修,CT球管,DR设备,医疗器械维修,医疗技术服务,医疗设备零备件

Error Code List

Error CodeMeaning
L8The examination room door is open.
H1When conducting radiography under the set X-ray conditions, the heat content exceeded the allowable heat content of the X-ray tube unit.
H2The heat content has exceeded the maximum heat capacity of the X-ray tube unit.
H4The thermal circuit of the X-ray tube unit has activated.
F90The temperature of main inverter IGBT is higher than 100 degC.
F64Radiography switch input error
F65Radiography preparation switch input error
ECPU memory battery replacement time
EEECPU memory battery expiration
F03-15V power supply control circuit fault
F11The measured tube voltage is at least 30 kV higher than the setting value.
F12The measured tube voltage is at least 70% lower than the setting value.
F13The measured tube voltage is approx. 10 kV or lower.
F14The X-ray tube has discharged. (3rd discharge)
F17 *The X-ray tube has discharged. (1st discharge)
F18 *The X-ray tube has discharged. (2nd discharge)
F19The tube voltage on either anode or cathode side is higher than 90 kV.
F21The measured tube current is at least 50 mA higher than the setting value.
F22The measured tube current is at least 50% lower than the setting value.
F23The measured tube current is less than approx. 8 mA.
F2FThe measured tube current is higher than approx. 1000 mA.
F31The measured filament current is at least 120% higher than the setting value.
F32The measured filament current is at least 80% lower than the setting value.
F33The measured filament current is approx. 1 A or lower.
F34The measured filament current is higher than the maximum focal spot current.
F41The measured capacitor voltage on the positive (+) side is higher than approx. 367 V.
F42The measured capacitor voltage on the positive (+) side is lower than approx. 235 V.
F43The measured capacitor voltage on the positive (+) side is lower than approx. 20 V.
F44The measured capacitor voltage on the negative (-) side is higher than approx. 367 V.
F45The measured capacitor voltage on the negative (-) side is lower than approx. 235 V.
F46The measured capacitor voltage on the negative (-) side is lower than approx. 20 V.
F55The measured capacitor voltage on the positive (+) side is at least 30 V higher than that on the negative (-) side.
F56The measured capacitor voltage on the negative (-) side is at least 30 V higher than that on the positive (+) side.
F61Starter fault (The starter does not activate normally at the start ofexposure. Or, the starter drive does not finish normally at the end of exposure.)
F62Bucky fault (No confirmation signal from the Bucky)
F6EExposure control circuit fault
F6FExposure control circuit fault
F6HExposure control circuit fault. Exposure interruption (no checking signal from FPD)
F71Dose area product communication timeout error
F72Dosimeter communication error
F73Dosimeter failure
F90IGBT device is in high temperature.
F91No additional filter is inserted though the additional filter setup has been done in the system.
F92An additional filter is inserted though the additional filter setup has not been done in the system.
F93The combination of the software installed (for emerging countries or for North America) and the software type memorized to EEPROM is notcorrect.Update to the software of correct combination or initialize the system.
F2aTube voltage feedback circuit fault
F1aTube voltage feedback circuit fault
F68FPD fault (No confirmation signal from FPD)
F76FPD radiography communication error
F80A tube voltage setting value out of the 40 kV to 150 kV range was transmitted from the DR.
F81A tube current setting value not set for any tube current position for RADspeed fit was transmitted from the DR.
F82A radiography time setting value out of the 1 msec - 10 msec range was transmitted from the DR.
F83An invalid tube focal spot was transmitted from the DR.
F84A tube current time product setting value out of the 0.5 mA to 500 mA range was transmitted from the DR.
F85An invalid phototimer pickup field was transmitted from the DR.
F86An invalid film density was transmitted from the DR.
F87An invalid film sensitivity was transmitted from the DR.
F88A program code out of the 0 to 432 range was transmitted from the DR.
F89An invalid technique code was transmitted from the DR.
F8aAn additonal filter number that cannot be selected was transmitted from the DR.